Sign In | Join Free | My ecer.jp
ecer.jp
Products
Search by Category
Home > Other Measuring & Analysing Instruments >

12m³ Walk-in Constant Temperature and Humidity Test Chamber IEC 60068-2-2

Categories Walk In Environmental Test Chamber
Brand Name: Sinuo
Model Number: SN886-12m³
Certification: Calibration Certificate (Cost Additional)
Place of Origin: China
MOQ: 1
Payment Terms: T/T
Supply Ability: 10 Sets Per Month
Delivery Time: 30 Days
Packaging Details: Export packing
Effective volume: 12m³
Temperature range: -40°~ 150℃
Humidity fluctuation: ±2.0%R.H
Humidity range: 20%RH~95%RH
Heating rate: ≦3.0℃/min (non-linear, no load)
Cooling rate: Average 1.0℃/min (non-linear, no load)
External dimensions Approx.: 3700 mm*2200 mm*2300 mm(W*H*D)
Inner box dimensions: 3000 mm*2000 mm*2000mm(W*H*D)
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

12m³ Walk-in Constant Temperature and Humidity Test Chamber IEC 60068-2-2


12m³ Walk-in Constant Temperature and Humidity Test Chamber IEC 60068-2-2

Complied Standards & Clauses



Complied Standard: IEC 60068-2-1, IEC 60068-2-2, IEC 60068-2-3, and IEC 60068-2-30,etc.

Product Overview


Application: This equipment is suitable for environmental simulation reliability tests such as high and low temperature constant and gradual changes, cycle tests, etc. for various electrical and electronic products and other products, components and materials.

Safety test limitations:

A) Testing and storage of flammable, explosive and volatile substance samples.

B) Testing of corrosive substance samples and testing or storage of biological samples.

C) Testing and storage of samples of strong electromagnetic emission sources.


Technical Parameters


Volume and dimensions
Effective volume12m³
Inner chamber size3000 mm*2000 mm*2000mm(W*H*D)

Outer chamber size

Approximately 3700 mm*2200 mm*2300 mm(W*H*D)
Warm reminder: Please refer to the three views of the final design for external dimensions.


Container size

The inner dimensions of the 40HQ container are 12.032m (L)*2.352m (W)*2.69m(H)
The door is 2.55M high (it will be extra wide after being packed with wooden boxes)

Performance
Cooling methodWater cooling
Temperature range-40°~ 150℃
Humidity range20%RH~95%RH
Temperature/humidity fluctuation±2℃/±2.0%RH
Temperature analysis accuracy0.01℃
Humidity analysis accuracy0.01RH
Temperature uniformity≤2℃
Humidity uniformity< 75%RH:±3%RH >75%RH±5%RH
Heating rate≦3.0℃/min (non-linear, no load)
Cooling rateAverage 1.0℃/min (non-linear, no load)
Load conditionNo load

Complied Standards

IEC 60068-2-1 Test A: Low-temperature test method.
IEC 60068-2-2 Test B: High-temperature test method.
IEC 60068-2-3 Test Ca: Constant damp heat test method.
IEC 60068-2-30 test Db: alternating damp heat test method.