Sign In | Join Free | My ecer.jp |
|
Categories | ETCO2 Sensor |
---|---|
Place of Origin: | China |
Brand Name: | Med Accessories |
Model Number: | MCMZ10 |
Packaging Details: | 1pcs/box |
Supply Ability: | 1000 Piece/Pieces per Month |
Certification: | CE,ISO,SFDA |
Delivery Time: | About one week |
Payment Terms: | T/T |
MOQ: | 1pcs |
Disinfecting Type: | Ultraviolet Light |
Properties: | Medical Materials & Accessories |
Size: | 3 meter |
Stock: | Yes |
Shelf Life: | 6months |
Material: | TPU |
Quality Certification: | CE/ISO13485 |
Instrument classification: | Class II |
Safety standard: | None |
Product name: | etCO2 sensor |
MOQ: | 1 Piece |
Port: | Shenzhen |
Company Info. |
Med Accessories Technology Dongguan Co., Ltd. |
Verified Supplier |
View Contact Details |
Product List |
Edan Mainstream etCO2 sensor CAPNOSTAT 5 compatible
ETCO2 Sensor Specification | |||||||||||||||
Project | Specifications | ||||||||||||||
Principle of Operation | Non-dispersive infrared(NDIR) single beam optics, dual wavelength, no moving parts. | ||||||||||||||
Initialization Time | Capnogram displayed in less than 3 seconds,full specifications within 3 minutes(At 25°C). | ||||||||||||||
CO2 Measurement Range | 0 - 114 mmHg; 0 - 15 %; 0 - 15.2 kPa. | ||||||||||||||
CO2 Rise Time | Rise Time Less than 80 ms - Adult Reusable or Single-Patient-Use
Airway Adapter Less than 80 ms - Infant Reusable or Single-Patient-Use Airway Adapter. | ||||||||||||||
CO2 Resolution | 0. 1mmHg ; 0 to 38 mmHg.; 0.25 mmHg ; 39 to 114mmHg. | ||||||||||||||
CO2 Accuracy | 0 - 40mmHg ±2 mmHg.; 41 - 76mmHg ±5% of reading.; 77 - 114 mmHg ±8%
of reading. Above 80 breaths per minute ±12% of reading. * Note: Gas emperature at 25°C. | ||||||||||||||
CO2 Stability | Short term drift: Drift over four hours shall not exceed 1 mmHg
maximum. Long term drift: Accuracy specification will be maintained over a 120-hour period. | ||||||||||||||
Sampling Frequency | 100Hz. | ||||||||||||||
Respiration Rate Range | 3 to 150 breaths per minute. | ||||||||||||||
Respiration Rate Accuracy | ±1 breath. | ||||||||||||||
Calibration | No routine user calibration required. An airway adapter zero is
required when changing to a different style of airway adapter. | ||||||||||||||
Airway Adapters | Single-patient-use or reusable, < 5 cc dead space (adult), < 1 cc dead space infant. | ||||||||||||||
Temperature | Operating: 0 to 40°C; Storage: -40 to 70°C; Transport: -40 to 70°C. | ||||||||||||||
Humidity | Operating: 10 to 90% RH, non-condensing; Storage: 10 to 90% RH,
non-condensing.; Transport: 10 to 90% RH, non-condensing. | ||||||||||||||
Atmospheric Pressure | Operating: 400-800 mmHg; Storage: 400-800 mmHg; Transport: 400-800 mmHg. | ||||||||||||||
Water Resistance | IPX4 - Splash-proof - Sensor only. | ||||||||||||||
Voltage Requirements | 5VDC ± 5 %. | ||||||||||||||
Power Rating | <1.0W Typical; <1.8 W Warm-up. | ||||||||||||||
Sensor Weight | <85g. | ||||||||||||||
Sensor Size | Deep*Wide*High:48mm * 24.3mm * 35.5mm. | ||||||||||||||
Cable Length | 2.4m. |
![]() |